Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 7-300: (1 year) CAL VER /5520 INSTRUMENT: Fluke 7-600: (1 year) CAL VER /5520 DATE: 22-Sep-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 25 NUMBER OF LINES: 103 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Fluke Interoffice Memo from Dan Parker to all service managers 8/1/96. # Subject: Fluke 7-300, 7-600, 12B 18 smf and 18/AL Calibration Procedures # # Compatibility: # 5520/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F 1.002 ASK+ K 1.003 HEAD EQUIPMENT SETUP 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP No warm-up time requirement. 1.005 DISP Ambient temperature: 18C - 28C. 1.005 DISP Relative humidity: less than 90%. 1.006 DISP Move switch to OFF position. 1.006 DISP 1.006 DISP Note: Erroneous result may occur if switch 1.006 DISP is not placed in OFF position before beginning tests. 1.007 HEAD {AC VOLTAGE TESTS} 1.008 DISP Move switch to the ON position. 1.009 HEAD AC VOLTAGE TESTS: {40V Range} 1.010 5520 40 4.40V 0.16U 50H SI 2W 2.001 5520 40 35.00V 1.05U 50H SI 2W 3.001 5520 40 35.00V 1.05U 400H SI 2W 4.001 JMPT 7.002 Fluke 7-600: (1 year) CAL VER /5520 4.002 HEAD AC VOLTAGE TESTS: {300V Range} 4.003 5520 300 100.0V 3.2U 50H SI 2W 5.001 5520 300 150.0V 4.6U 400H SI 2W 6.001 5520 300 300.0V 9.0U 50H SI 2W 7.001 JMP 11.001 7.002 HEAD AC VOLTAGE TESTS: {400V Range} 7.003 5520 400 100.0V 3.2U 50H SI 2W 8.001 5520 400 150.0V 4.6U 400H SI 2W 9.001 5520 100.0V 80H SI S 2W 9.002 PORT [D1000] 9.003 5520 150.0V 80H SI S 2W 9.004 PORT [D1000] 9.005 5520 200.0V 80H SI S 2W 9.006 PORT [D1000] 9.007 5520 250.0V 80H SI S 2W 9.008 PORT [D1000] 9.009 5520 300.0V 80H SI S 2W 9.010 PORT [D1000] 9.011 5520 400 350.0V 10.5U 80H SI 2W 10.001 HEAD AC VOLTAGE TESTS: {600V Range} 10.002 5520 600 600V 20U 400H SI 2W 11.001 HEAD {DC VOLTAGE TESTS} 11.002 HEAD DC VOLTAGE TESTS: {40V Range} 11.003 5520 40 4.40V 0.08U 2W 12.001 5520 40 -35.00V 0.54U 2W 13.001 JMPT 16.002 Fluke 7-600: (1 year) CAL VER /5520 13.002 HEAD DC VOLTAGE TESTS: {300V Range} 13.003 5520 300 100.0V 1.6U 2W 14.001 5520 300 -150.0V 2.3U 2W 15.001 5520 300 300.0V 4.6U 2W 16.001 JMP 20.001 16.002 HEAD DC VOLTAGE TESTS: {400V Range} 16.003 5520 400 100.0V 1.6U 2W 17.001 5520 400 -150.0V 2.3U 2W 18.001 5520 400 300.0V 4.6U 2W 19.001 HEAD DC VOLTAGE TESTS: {600V Range} 19.002 5520 600 -600V 10U 2W 20.001 HEAD {RESISTANCE TESTS} 20.002 HEAD RESISTANCE TESTS: {400 Ohm Range} 20.003 5520 400 0.0Z +0.2U 2W 21.001 5520 400 1.0Z 0.2U 2W 22.001 5520 400 350.0Z 5.5U 2W 23.001 HEAD {CONTINUITY BEEPER} 23.002 5520 0.0Z S 2W 23.003 EVAL -s "0 Ohms" : Is beeper on? 24.001 5520 1.0Z S 2W 24.002 EVAL -s "1 Ohm" : Is beeper on? 25.001 END